Knowledge Base Feature

Knowledge Base
Home / Knowledge Base

Hettich Knowledge Base



Results...

Knowledge Base

Category: Symbols & Markings

Topic: Symbols

Article Title: Disposal of Centrifuge

Before disposal, the centrifuge must be decontaminated and free of any bio- or radioactive hazard to protect people, the environment and property.

Category: Symbols & Markings
 Topic: Symbols
Article Title: Symbols used on Centrifuge and in Ops Manual

Category: Operation & Use
Topic: Centrifuge Operation
Article Title: Tacho Errors 1, 2, 96; -1- (EBA 20, EBA 20S, EBA 270, ROTOFIX 32_A only)


Tacho (tachometer) errors are errors in the system that recognizes motion of the rotor.

To clear Tacho Errors 1, 2, & 96:

Open lid, turn power off. Make sure rotor is properly attached to the motor.

ASK A QUESTION