Category: Quality & Safety
Topic: Centrifugation Safety
Article Title: Approved Rotors and Accessories
Hettich manufactures a variety of rotors, buckets and accessories for use in their centrifuges.
Category: Care & Maintenance
Topic: Decontamination
Article Title: Autoclaving
Most facilities have a pre-determined schedule of maintenance for the different categories of equipment. Hettich recommends all centrifuges be cleaned and inspected at "regular intervals". For high usage (24 hours a day/ 7 days a week), once a month is strongly suggested. For mid to low usage, (normal business hours, weekdays only), every three months.
Category: Quality & Safety
Topic: Centrifugation Safety
Article Title: Bio Safety
Hettich manufacturers a variety of buckets and lids designed for bio-containment. These lids and buckets have a special screw cap design with bio-safety seal that contains any accidental tube breakage and traps it within the bucket.
Category: Care & Maintenance
Topic: Decontamination
Article Title: Decontamination Certificate
For the safety of transportation personnel as well as our own receiving department, a certificate of decontamination MUST accompany all products returned to Hettich Instruments.
Category: Operation & Use
Topic: Centrifuge Operation
Article Title: Changing Centrifugation Parameters During a Run
The running time, the speed, the relative centrifugal force (RCF), the centrifuging radius, the run-up and run-down parameters as well as the temperature (for centrifuges with cooling only) can all be changed while the centrifuge is running.
Category: Care & Maintenance
Topic: Preventative Maintenance
Article Title: Clean, Inspect & Lubricate
It is important to regularly clean, inspect and lubricate the rotor and accessories in all Hettich centrifuges. General maintenance procedures will ensure proper operations and extend the life span of all components, reduce the operational noise, decrease potential vibrations and prevent imbalance errors from occurring.
Category: Operation & Use
Topic: Centrifuge Operation
Article Title: Continuous Operation
Set the hours, minutes, and seconds “0” or “—“. Press the START key.
Category: Quality & Safety
Topic: Centrifugation Safety
Article Title: Corrosive Agents
Hettich manufactures its products using high quality materials that are engineered to last for many years.
Category: Operation & Use
Topic: Centrifuge Operation
Article Title: Definition of Mains Reset
A mains reset is the removal and re-application of AC power via the power switch.
Category: Care & Maintenance
Topic: Decontamination
Article Title: Disinfection of rotors and accessories
If infectious material should get on Hettich rotors or accessories, they must be appropriately disinfected. According to the operator’s manual, these are the most desirable ingredients of suitable disinfectants: